Buy me a coffee
Home › FDA recalls › Evolution upgrade 1.5T amd Evolution upgrade 3.0T
CLASS II · ModerateFDA

Philips North America recalls Evolution upgrade 1.5T amd Evolution upgrade 3.0T

📅 Aug 28, 2026 · 📍 Nationwide
Recalling company
Philips North America
Product
Evolution upgrade 1.5T amd Evolution upgrade 3.0T; Model number (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782166 (1.5T), 4. 782201(1.5T), 5. 782117 (3.0T), 6. 782143 (3.0T), 7. 782162 (3.0T); Software Versions: R11, R12 and R13;
Reason
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Aug 28, 2026
Reported
Oct 7, 2026
Distribution
Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
Status
Ongoing
Quantity
91
Lot / codes
1. Model number (REF): 782116 (1.5T); UDI-DI: 00884838099722; Serial Numbers: 41042, 41235, 41287, 41295, 41367, 41505, 41568, 70068, 70093, 70995, 70996, 84370; 2. Model number (REF): 782148 (1.5T); UDI-DI: 00884838108714; Serial Numbers: 41011, 41031, 41110, 41244, 41277, 41334, 41427, 41446, 41464, 41523, 41534, 70425, 70454, 70473, 70622, 70630, 70644, 70769, 70803, 70966, 70983, 84102, 84409, 84560, 84569, 84630, 84825; 3. Model number (REF): 782166 (1.5T); UDI-DI: 00884838115774; Serial Numbers: 41005, 41161, 41485, 41499, 41572, 70155, 70206, 70826, 84048, 84270, 84541; 4. Model number (REF): 782201 (1.5T); UDI-DI: 00884838128552; Serial Numbers: 84018, 84648; 5. Model number (REF): 782117 (3.0T); UDI-DI: 00884838099739; Serial Numbers: 42023, 42024, 42078, 42375, 71348, 85681, 85705; 6. Model number (REF): 782143 (3.0T); UDI-DI: 00884838108660; Serial Numbers: 42069, 42225, 42254, 42318, 42366, 42415, 42440, 71389, 71460, 71595, 71815, 71836, 85649, 85651, 85654, 85682, 85683, 85688, 85690, 85695, 85697, 85699, 85701, 85714; 7. Model number (REF): 782162 (3.0T); UDI-DI: 00884838115736; Serial Numbers: 42184, 42369, 71074, 71403, 71433, 71574, 71740, 85145;
Company location
Cambridge, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-3318-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3318-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.