Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography System
- Recalling company
- Shanghai United Imaging Healthcare Co., Ltd.
- Product
- Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama GS with software
- Reason
- Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Aug 11, 2026
- Reported
- Oct 7, 2026
- Distribution
- US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.
- Status
- Ongoing
- Quantity
- 3
- Lot / codes
- UDI-DI: 06971576832187. Serial Number: 11MIP00007/R001.17.5.1785268, 11MIP00009/R001.18.2.1967222, 11MIP00010/R001.18.2.1967222
- Company location
- Shanghai
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-3285-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-3285-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.