Buy me a coffee
Home › FDA recalls › Fisher Scientific Sure-Vue Serum/Urine hCG
CLASS II · ModerateFDA

Alere San Diego, Inc. recalls Fisher Scientific Sure-Vue Serum/Urine hCG

📅 Aug 20, 2026 · 📍 Nationwide
Recalling company
Alere San Diego, Inc.
Product
Fisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman Coulter Inc's ICON 25 hCG, Model/ Part: FHC-202/43025; Henry Schein One Step+ hCG Combo test in Serum or Urine, Model/ Part: FHC-202/9004075; Alere hCG Combo Cassette (25mIU/mL) in Urine (CLIA Waived) or Serum, Model/ Part: FHC-202/92225; Pro Advantage By NDC Urine/Serum Pregnancy Cassette Device, Model/ Part: FHC-202/P080021
Reason
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Aug 20, 2026
Reported
Oct 7, 2026
Distribution
Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden.
Status
Ongoing
Quantity
5,331,585
Lot / codes
All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900529/00613647000540/1155810906, 2018-01-15; 23900528/00613647000564/1155810906, 2018-01-15; 43025/15099590776886/May 2015; 9004075/10304040024123/1155875804, 2014-06-09; 92225/10815845020116/1156015603, 07/16; P080021/M220P0800213/40001856, 2023-03-30
Company location
San Diego, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-3272-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3272-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.