Buy me a coffee
Home › FDA recalls › Automated Impella Controller
CLASS II · ModerateFDA

Abiomed, Inc. recalls Automated Impella Controller

📅 Feb 19, 2019 · 📍 Nationwide
Recalling company
Abiomed, Inc.
Product
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.
Reason
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Feb 19, 2019
Reported
Sep 30, 2026
Distribution
Worldwide - US Nationwide and the countries of Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, United Kingdom.
Status
Ongoing
Quantity
3,322 units
Lot / codes
1. Product Code: 0042-0000-CA. GTIN: 00813502011272. Serial Number: IC2687. 2. Product Code: 0042-0000-EU. GTIN: 00813502011289. Serial Numbers: IC3875, IC2380, IC2331, IC1721, IC1593, IC1587, IC4911, IC4231, IC4232, IC6844, IC2221, IC4628, IC6474, IC4502, IC1398, IC4055, IC3816, IC6810, IC6812, IC6449, IC6469, IC6441, IC6443, IC6440, IC6436, IC6437, IC6439, IC6272, IC6149, IC6081, IC5681, IC5682, IC5517, IC5177, IC5169, IC5086, IC4892, IC4852, IC4848, IC4875, IC4851, IC4869, IC4823, IC4824, IC4821, IC4822, IC4845, IC4840, IC4819, IC4820, IC4817, IC4818, IC4815, IC4816, IC4796, IC4797, IC4793, IC4795, IC4792, IC4794, IC4763, IC4781, IC4762, IC4760, IC4761, IC4758, IC4759, IC4748, IC4757, IC4737, IC4738, IC4735, IC4736, IC4733, IC4734, IC4678, IC4677, IC4646, IC4632, IC4633, IC4630, IC4631, IC4585, IC4629, IC4626, IC4613, IC4614, IC4568, IC4610, IC4612, IC4557, IC4567, IC4594, IC4595, IC4559, IC4564, IC4560, IC4561, IC4505, IC4506, IC4500, IC4469, IC4466, IC4467, IC4468, IC4446, IC4442, IC4443, IC4444, IC4445, IC4370, IC4371, IC4367, IC4365, IC4366, IC4361, IC4328, IC4234, IC4363, IC4362, IC4356, IC4357, IC4322, IC4258, IC4252, IC4253, IC4251, IC4242, IC4240, IC4241, IC4235, IC4238, IC4236, IC4230, IC4227, IC4228, IC4229, IC4210, IC4212, IC4207, IC4208, IC4217, IC4148, IC4149, IC4146, IC4147, IC4128, IC4129, IC4096, IC4097, IC4098, IC4083, IC4036, IC4037, IC3889, IC3833, IC3834, IC3859, IC3861, IC3862, IC3863, IC3885, IC3886, IC3884, IC3871, IC3872, IC3873, IC3832, IC3860, I
Company location
Danvers, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-3251-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3251-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.