ICU Medical, Inc. recalls CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-760
- Recalling company
- ICU Medical, Inc.
- Product
- CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24JP 21-7600-24JP 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24
- Reason
- Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Aug 20, 2026
- Reported
- Sep 23, 2026
- Distribution
- Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, LA, MI, MO, MS, TN, TX, VA, WI and the countries of Austria, Australia, Belgium, Costa Rica, Germany, Estonia, Spain, Finland, France, Great Britain, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, and Sweden.
- Status
- Ongoing
- Quantity
- 185,688 units
- Lot / codes
- 21-7600-24 UDI-DI 30610586044001 Lots 6026878 6037784 6054015 6062724 6070616 6082332 6092886 6101906 6108556 6126029 6140893 6143454 6147696 6147697 6162769 6155095 6166032 6171587 6171586 21-7609-24 UDI-DI 30610586044018 Lot 6022079 6037697 6037698 6053978 6061525 6072970 6072964 6070575 6077773 6092846 6101798 6108514 6108515 6108516 6140862 6140861 6140865 6155132
- Company location
- Lake Forest, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-3212-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-3212-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.