Buy me a coffee
Home › FDA recalls › TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaed
CLASS II · ModerateFDA

Howmedica Osteonics Corp. recalls TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufact

📅 Jul 30, 2026 · 📍 Nationwide
Recalling company
Howmedica Osteonics Corp.
Product
TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jul 30, 2026
Reported
Sep 23, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Status
Ongoing
Quantity
12,156 US
Lot / codes
620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0
Company location
Mahwah, NJ
Recall type
Voluntary: Firm initiated
Recall number
Z-3204-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3204-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.