Buy me a coffee
Home › FDA recalls › TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIA
CLASS II · ModerateFDA

Howmedica Osteonics Corp. recalls TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM

📅 Jul 30, 2026 · 📍 Nationwide
Recalling company
Howmedica Osteonics Corp.
Product
TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jul 30, 2026
Reported
Sep 23, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Status
Ongoing
Quantity
12,156 US
Lot / codes
5612-P-411 UDI-DI 07613327601893 Lot 360958 5612-P-413 UDI-DI 07613327602012 Lot 354573
Company location
Mahwah, NJ
Recall type
Voluntary: Firm initiated
Recall number
Z-3198-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3198-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.