Buy me a coffee
Home › FDA recalls › Aisys CS2 with Et Control
CLASS II · ModerateFDA

Datex-Ohmeda, Inc. recalls Aisys CS2 with Et Control

📅 Jul 30, 2026 · 📍 Nationwide
Recalling company
Datex-Ohmeda, Inc.
Product
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
Reason
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jul 30, 2026
Reported
Sep 23, 2026
Distribution
Worldwide - US Nationwide distribution.
Status
Ongoing
Quantity
7 units
Lot / codes
Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
Company location
Madison, WI
Recall type
Voluntary: Firm initiated
Recall number
Z-3191-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3191-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.