Buy me a coffee
Home › FDA recalls › Brand Name: VERIQA
CLASS II · ModerateFDA

PTW FREIBURG recalls Brand Name: VERIQA

📅 Aug 4, 2026 · 📍 Nationwide
Recalling company
PTW FREIBURG
Product
Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;
Reason
When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Aug 4, 2026
Reported
Sep 23, 2026
Distribution
Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, Portugal, Romania, T¿rkiye, Chile, United Arab Emirates, Saudi Arabia, Mexico, Hong Kong, Brazil, Tajikistan, Ecuador, Iran, Thailand, Serbia, South Africa, Kazakhstan, Algeria.
Status
Ongoing
Quantity
2
Lot / codes
Model/Catalog Number: S070031.520; UDI-DI: +EPTWS070031Q5200; Software Version 3.1.18;
Company location
Freiburg Im Breisgau, N/A
Recall type
Voluntary: Firm initiated
Recall number
Z-3174-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3174-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.