Buy me a coffee
Home › FDA recalls › StatStrip Glucose Hospital Meter System.
CLASS I · SeriousFDA

Nova Biomedical Corporation recalls StatStrip Glucose Hospital Meter System.

📅 Jul 29, 2026 · 📍 Nationwide
Recalling company
Nova Biomedical Corporation
Product
StatStrip Glucose Hospital Meter System. Model No: 66739. Glucose test system.
Reason
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Jul 29, 2026
Reported
Sep 23, 2026
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.
Status
Ongoing
Quantity
10,941 units
Lot / codes
Model No: 66739. UDI-DI: 0038548063685. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
Company location
Waltham, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-3129-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3129-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.