GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.1
- Recalling company
- GE Medical Systems Information Technologies Inc
- Product
- PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209
- Reason
- Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jul 20, 2026
- Reported
- Sep 2, 2026
- Distribution
- US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.
- Status
- Ongoing
- Quantity
- 23 units
- Lot / codes
- GTIN 00195278679284, Serial Numbers: SWT25250001TA, SWT25360001TA, SWT25360002TA, SWT24090006TA, SWT25100001TA, SWT25120002TA, SWT24140007TA, SWT24150008TA, SWT25100004TA, SWT24050001TA, SWT25110003TA, SWT24180011TA, SWT25120003TA, SWT25230001TA, SWT25230002TA, SWT25110009TA, SWT25110010TA, SWT25230003TA, SWT25230005TA, SWT25230006TA, SWT25230004TA, SWT25260001TA, SWT31240012TA.
- Company location
- Waukesha, WI
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-3036-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-3036-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.