Buy me a coffee
Home › FDA recalls › ORT300
CLASS II · ModerateFDA

IMRIS Imaging Inc recalls ORT300

📅 Aug 6, 2026 · 📍 Nationwide
Recalling company
IMRIS Imaging Inc
Product
ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600
Reason
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Aug 6, 2026
Reported
Sep 2, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
Status
Ongoing
Quantity
13 units
Lot / codes
114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376
Company location
Chaska, MN
Recall type
Voluntary: Firm initiated
Recall number
Z-3034-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-3034-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.