Buy me a coffee
Home › FDA recalls › Brand Name: Veritas Standard Reverse Total Shoulder Product
CLASS II · ModerateFDA

restor3d Inc. recalls Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoi

📅 Jun 30, 2026 · 📍 Nationwide
Recalling company
restor3d Inc.
Product
Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder
Reason
Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jun 30, 2026
Reported
Aug 26, 2026
Distribution
U.S. Nationwide distribution in the state of Texas.
Status
Ongoing
Quantity
59
Lot / codes
Lot Code: Model No 1500-1030; UDI-DI 00840097513225; Lot Numbers A00030, A00065, A09828, A09830 Model No 1500-1035; UDI-DI 00840097513232; Lot Numbers A00064, A09829
Company location
Durham, NC
Recall type
Voluntary: Firm initiated
Recall number
Z-2987-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2987-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.