Medtronic MiniMed, Inc. recalls MiniMed 640G Insulin Pump
- Recalling company
- Medtronic MiniMed, Inc.
- Product
- MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
- Reason
- Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jul 13, 2026
- Reported
- Aug 26, 2026
- Distribution
- International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
- Status
- Ongoing
- Quantity
- 16
- Lot / codes
- Model/UDI-DI/Serial Numbers: MMT-1711/00763000404550/NG2660048H, NG2663343H, NG3015306H; MMT-1712/00763000404567/NG2628486H, NG2758213H, NG2999651H, NG3349198H, NG3351719H, NG2678842H, NG2831821H, NG2831546H, NG2932027H, NG3049657H, NG3033094H, NG2831538H, NG3738739H
- Company location
- Northridge, CA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2980-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2980-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.