Beckman Coulter Inc. recalls Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower:
- Recalling company
- Beckman Coulter Inc.
- Product
- Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers
- Reason
- a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jul 15, 2026
- Reported
- Aug 26, 2026
- Distribution
- Worldwide distribution - US Nationwide.
- Status
- Ongoing
- Lot / codes
- 1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
- Company location
- Brea, CA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2974-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2974-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.