Buy me a coffee
Home › FDA recalls › Beckman Coulter DxC 500i Clinical Analyzer Modules using sof
CLASS II · ModerateFDA

Beckman Coulter Inc. recalls Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower:

📅 Jul 15, 2026 · 📍 Nationwide
Recalling company
Beckman Coulter Inc.
Product
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers
Reason
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jul 15, 2026
Reported
Aug 26, 2026
Distribution
Worldwide distribution - US Nationwide.
Status
Ongoing
Lot / codes
1) Part Number C63522, UDI/DI 14987666545089; 2) Part Number C63521, UDI/DI 14987666545072; 3) Part Number C13252, UDI/DI 15099590742331.
Company location
Brea, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-2973-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2973-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.