Buy me a coffee
Home › FDA recalls › TD Navio Midi Speech Generating Device
CLASS II · ModerateFDA

Tobii Dynavox Llc recalls TD Navio Midi Speech Generating Device

📅 Jul 8, 2026 · 📍 Nationwide
Recalling company
Tobii Dynavox Llc
Product
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Reason
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jul 8, 2026
Reported
Aug 26, 2026
Distribution
US Nationwide distribution.
Status
Ongoing
Lot / codes
UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).
Company location
Pittsburgh, PA
Recall type
Voluntary: Firm initiated
Recall number
Z-2959-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2959-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.