Arrowhead DE LLC recalls Arrow-LOK Implant
- Recalling company
- Arrowhead DE LLC
- Product
- Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550, 21-353513, 21-353516, 27-253519, 27-253522, 27-253525, 27-253528, 27-253530, 27-253533, 27-253535. Implant indicated for fixation of osteotomies, arthrodeses and reconstruction in the lesser toes following corrective procedures.
- Reason
- Implant did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jun 23, 2026
- Reported
- Aug 26, 2026
- Distribution
- US Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.
- Status
- Ongoing
- Quantity
- 4,171
- Lot / codes
- REF/UDI-DI/Lot: 21-253519/00855967006035/24662119, 242162; 21-253522/00855967006042/24662122, 242163, 220702; 21-253525/00855967006059/24662125, 220703, 242164; 21-253528/00855967006066/24662128, 242165, 230704; 21-253530/00855967006073/242166, 230705, 220705; 21-253533/00855967006080/24662133, 242167, 230706; 21-253535/00855967006097/24662135, 242168, 230707; 21-253540/00855967006110/242169; 21-253545/00855967006127/242170; 21-253550/00855967006134/24662150; 21-353513/00855967006004/24662113, 242160; 21-353516/00855967006011/24662116, 230701, 242161; 27-253519/00855967006028/24662719, 242171; 27-253522/00855967006141/242172; 27-253525/00855967006158/242173; 27-253528/00855967006165/242174; 27-253530/00855967006172/242175; 27-253533/00855967006189/24662733, 242176; 27-253535/00855967006196/24662735, 242177
- Company location
- Peachtree City, GA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2958-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2958-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.