Buy me a coffee
Home › FDA recalls › ProNephro AKI
CLASS II · ModerateFDA

Bioporto Diagnostics A/S recalls ProNephro AKI

📅 May 18, 2026 · 📍 See notice
Recalling company
Bioporto Diagnostics A/S
Product
ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.
Reason
The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
May 18, 2026
Reported
Aug 19, 2026
Distribution
Worldwide - The affected product distribution included the United States and the countries of European Union, Chile, and Hong Kong. No affected product was distributed to Canada; however, Health Canada was notified because the product is registered in Canada.
Status
Ongoing
Quantity
400 units
Lot / codes
Model number: ST101UB / 08787174 190; GTIN code: 5711435000156; Lot-number: 04909125 and 04912125.
Company location
Hellerup
Recall type
Voluntary: Firm initiated
Recall number
Z-2937-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2937-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.