Buy me a coffee
Home › FDA recalls › Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT
CLASS II · ModerateFDA

TISSIUM SA recalls Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT

📅 Jun 15, 2026 · 📍 Nationwide
Recalling company
TISSIUM SA
Product
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Reason
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jun 15, 2026
Reported
Aug 12, 2026
Distribution
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Status
Ongoing
Lot / codes
Lot Code: Model No NCOCO2010; UDI-DI 03760279380094; Lot Number 250700042, 251000033, and 251000034
Company location
Paris 12e Arrondissement
Recall type
Voluntary: Firm initiated
Recall number
Z-2903-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2903-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.