Buy me a coffee
Home › FDA recalls › REF # D23352
CLASS II · ModerateFDA

Beckman Coulter Inc. recalls REF # D23352

📅 Jun 26, 2026 · 📍 Nationwide
Recalling company
Beckman Coulter Inc.
Product
REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument
Reason
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jun 26, 2026
Reported
Aug 12, 2026
Distribution
US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.
Status
Ongoing
Quantity
158 instruments
Lot / codes
UDI: 15099590799243/ V5.03 Panel Update 11
Company location
Brea, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-2888-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2888-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.