Hardy Diagnostics recalls HardyDisk AST Cefiderocol 30¿g

📅 Jun 8, 2026 · 📍 Nationwide
Recalling company
Hardy Diagnostics
Product
HardyDisk AST Cefiderocol 30¿g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30¿g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Pseudomonas aeruginosa, Acinetobacter baumannii complex and Stenotrophomonas maltophilia to Cefiderocol,
Reason
Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jun 8, 2026
Reported
Aug 12, 2026
Distribution
U.S. Nationwide distribution in the states of AZ, CA, FL, ID, IN, MA, MS, NJ, NV, NY, OH, PA, and TN.
Status
Ongoing
Quantity
43 units
Lot / codes
Catalog Number: Z9435 UD-DI code: Pending Lot Number: 543461
Company location
Santa Maria, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-2887-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2887-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.