Buy me a coffee
Home › FDA recalls › INFINION CX LEAD KIT
CLASS I · SeriousFDA

Boston Scientific Neuromodulation Corporation recalls INFINION CX LEAD KIT

📅 Jun 17, 2026 · 📍 NY
Recalling company
Boston Scientific Neuromodulation Corporation
Product
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.
Reason
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Jun 17, 2026
Reported
Aug 12, 2026
Distribution
US: NY. OUS: Canada, Portugal, Spain, France, Netherlands, Great Britain, Italy, Israel, Serbia, Greece, Germany, United Arab Emirates, Ireland, Hungary, Czech Republic, Brazil, Colombia, Argentina
Status
Ongoing
Quantity
1079
Lot / codes
Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products.
Company location
Valencia, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-2879-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2879-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.