Buy me a coffee
Home › FDA recalls › Nerivio Infinity.
CLASS II · ModerateFDA

THERANICA BIOELECTRONICS LTD recalls Nerivio Infinity.

📅 Mar 23, 2026 · 📍 See notice
Recalling company
THERANICA BIOELECTRONICS LTD
Product
Nerivio Infinity. Model/Catalog Number: FGD000100. Software Version: FW Version 4.1.2.0, App version 7.28.0.11
Reason
Potential for the batter to have a thermal-runway event during charging, resulting in smoke and odor.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Mar 23, 2026
Reported
Aug 12, 2026
Distribution
The product's entry point to the US is a 3PL warehouse from which it is being shipped to DME distributors or specialty pharmacy who perform the shipment to the end-user
Status
Ongoing
Quantity
11,886 units
Lot / codes
Model/Catalog Number: FGD000100. UDI: 01072900175781181125120221NI2544-0300673 Software Version: FW Version 4.1.2.0, App version 7.28.0.11 GTIN: 7290017578118. Serial Number: NI2544-0300673.
Company location
Netanya
Recall type
Voluntary: Firm initiated
Recall number
Z-2855-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2855-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.