Buy me a coffee
Home › FDA recalls › EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR
CLASS II · ModerateFDA

Medtronic Perfusion Systems recalls EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR

📅 Jun 24, 2026 · 📍 Nationwide
Recalling company
Medtronic Perfusion Systems
Product
EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622
Reason
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jun 24, 2026
Reported
Aug 5, 2026
Distribution
Worldwide - US Nationwide distribution.
Status
Ongoing
Quantity
13,510 units
Lot / codes
UDI-DI: 20763000135670, Lot Numbers: 0232601501 0232601510 0232601549 0232601550 0232602178 0232602290 0232602302 0232910952 0232910953 0232910954 0232910955 0233244075 0233244301 0233244531 0233286362 0233286374 0233286390 0233409034 0233416071 0233528731
Company location
Brooklyn Park, MN
Recall type
Voluntary: Firm initiated
Recall number
Z-2808-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2808-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.