Buy me a coffee
Home › FDA recalls › M5 Corpus Power Wheelchair REF
CLASS II · ModerateFDA

Permobil Inc recalls M5 Corpus Power Wheelchair REF

📅 Jun 19, 2025 · 📍 Nationwide
Recalling company
Permobil Inc
Product
M5 Corpus Power Wheelchair REF,120032
Reason
This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jun 19, 2025
Reported
Aug 5, 2026
Distribution
Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.
Status
Ongoing
Quantity
715 wheelchairs
Lot / codes
Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562
Company location
Lebanon, TN
Recall type
Voluntary: Firm initiated
Recall number
Z-2805-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2805-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.