Medline Industries, LP recalls Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters
- Recalling company
- Medline Industries, LP
- Product
- Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. PROSTATE ROBOTIC DYNJ901822B
- Reason
- Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Jun 8, 2026
- Reported
- Jul 29, 2026
- Distribution
- US Nationwide distribution in the states of CA, FL, MD. PA, and TX.
- Status
- Ongoing
- Quantity
- 980 kits
- Lot / codes
- DYNJ901822B UDI-DI 10195327319182 Lots 25KBA955 26CBL889
- Company location
- Northfield, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2783-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2783-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.