Buy me a coffee
Home › FDA recalls › SnapshotGLO Wound Imaging Device
CLASS II · ModerateFDA

KENT IMAGING, INC. recalls SnapshotGLO Wound Imaging Device

📅 Jun 30, 2026 · 📍 Nationwide
Recalling company
KENT IMAGING, INC.
Product
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
Reason
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jun 30, 2026
Reported
Jul 29, 2026
Distribution
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
Status
Ongoing
Quantity
77
Lot / codes
UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
Company location
Calgary, N/A
Recall type
Voluntary: Firm initiated
Recall number
Z-2775-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2775-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.