PIE Medical Imaging B.V. recalls 3mensio Workstation
- Recalling company
- PIE Medical Imaging B.V.
- Product
- 3mensio Workstation (Vascular Fenestrated) software
- Reason
- User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- May 7, 2026
- Reported
- Jul 29, 2026
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.
- Status
- Ongoing
- Quantity
- 165
- Lot / codes
- Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5
- Company location
- Maastricht, N/A
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2767-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2767-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.