Buy me a coffee
Home › FDA recalls › 3mensio Workstation
CLASS II · ModerateFDA

PIE Medical Imaging B.V. recalls 3mensio Workstation

📅 May 7, 2026 · 📍 Nationwide
Recalling company
PIE Medical Imaging B.V.
Product
3mensio Workstation (Vascular Fenestrated) software
Reason
User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
May 7, 2026
Reported
Jul 29, 2026
Distribution
Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.
Status
Ongoing
Quantity
165
Lot / codes
Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5
Company location
Maastricht, N/A
Recall type
Voluntary: Firm initiated
Recall number
Z-2767-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2767-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.