Home › FDA recalls › Baxter Duo-Vent Solution Set and Male Luer Lock Adapter
CLASS I · SeriousFDA

Baxter Healthcare Corporation recalls Baxter Duo-Vent Solution Set and Male Luer Lock Adapter

📅 Jun 30, 2026 · 📍 Nationwide
Recalling company
Baxter Healthcare Corporation
Product
Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets
Reason
Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Jun 30, 2026
Reported
Aug 19, 2026
Distribution
US Nationwide. No OUS consignees.
Status
Ongoing
Quantity
24528 units
Lot / codes
UDI/DI 00085412014319, Lot Numbers: DR26C27062, exp. 30-MAR-2028; DR26D14031, exp. 15-Apr-2028; DR26E13055, exp. 14-May-2028
Company location
Deerfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-2766-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2766-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.