Buy me a coffee
Home › FDA recalls › Photera 600 3-D
CLASS II · ModerateFDA

Daavlin Distributing Company recalls Photera 600 3-D

📅 Mar 5, 2026 · 📍 Nationwide
Recalling company
Daavlin Distributing Company
Product
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Reason
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Mar 5, 2026
Reported
Jul 22, 2026
Distribution
US Nationwide distribution.
Status
Ongoing
Quantity
212 units
Lot / codes
Serial Numbers: 6-09278NB8CL, 6-09271NB8CL, 6-09055NB8CL, 6-09113NB8CL, 6-09204NB8CL, 6-09207NB8CL, 6-09116NB8CL, 6-09263NB8CL, 6-09272NB8CL, 6-09118NB8CL, 6-09275NB8CL, 6-09264NB8CL, 6-09060NB8CL, 6-09122NB8CL, 6-09216NB8CL, 6-09125NB8CL, 6-09353NB8CL, 6-09346NB8CL, 6-09128NB8CL, 6-09129NB8CL, 6-09133NB8CL, 6-09134NB8CL, 6-09266NB8CL, 6-09215NB8CL, 6-09355NB8CL, 6-09137NB8CL, 6-09359NB8CL, 6-09343NB8CL, 6-09140NB8CL, 6-09142NB8CL, 6-09143NB8CL, 6-09273NB8CL, 6-09065NB8CL, 6-09144NB8CL, 6-09237NB8CL, 6-09225NB8CL, 6-09145NB8CL, 6-09147NB8CL, 6-09148NB8CL, 6-09378NB8CL, 6-09252NB8CL, 6-09051NB8CL, 6-09260NB8CL, 6-09250NB8CL, 6-09149NB8CL, 6-09238NB8CL, 6-09212NB8CL, 6-09277NB8CL, 6-09048NB8CL, 6-09274NB8CL, 6-09259NB8CL, 6-09224NB8CL, 6-09057NB8CL, 6-09436NB12CL, 6-09438NB8CL, 6-09434NB12CL, 6-09270NB8CL, 6-09115NB8CL, 6-09111NB8CL, 6-09435NB12CL, 6-09124NB8CL, 6-09332NB12CL, 6-09223NB8CL, 6-09219NB8CL, 6-09234NB8CL, 6-09063NB8CL, 6-09232NB8CL, 6-09479NB12CL, 6-09373NB8CL, 6-09127NB8CL, 6-09247NB8CL, 6-09135NB8CL, 6-09344NB8CL, 6-09221NB8CL, 6-09050NB8CL, 6-09336NB12CL, 6-09049NB8CL, 6-09068NB8CL, 6-09331NB12CL, 6-09240NB8CL, 6-09235NB8CL, 6-09365NB8CL, 6-09205NB8CL, 6-09345NB8CL, 6-09202NB8CL, 6-09136NB8CL, 6-09201NB8CL, 6-09249NB8CL, 6-09437NB8CL, 6-09208NB8CL, 6-09434NB12CL, 6-09503NB8CL, 6-09132NB8CL, 6-09228NB8CL, 6-09257NB8CL, 6-09058NB8CL, 6-09210NB8CL, 6-09054NB8CL, 6-09062NB8CL, 6-09254NB8CL, 6-09256NB8CL, 6-09255NB8CL, 6-09229NB8CL, 6-09062NB8CL, 6-09203NB8CL, 6-093
Company location
Bryan, OH
Recall type
Voluntary: Firm initiated
Recall number
Z-2744-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2744-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.