Buy me a coffee
Home › FDA recalls › MiniMed 780G Insulin Pump
CLASS II · ModerateFDA

Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump

📅 May 27, 2026 · 📍 See notice
Recalling company
Medtronic MiniMed, Inc.
Product
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Reason
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
May 27, 2026
Reported
Jul 15, 2026
Distribution
International distribution to the countries of Spain, Great Britain, Netherlands, Italy.
Status
Ongoing
Quantity
778
Lot / codes
UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62
Company location
Northridge, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-2698-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2698-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.