Buy me a coffee
Home › FDA recalls › Philips Telemetry Monitor 5500 1.4 GHz.
CLASS II · ModerateFDA

Philips North America Llc recalls Philips Telemetry Monitor 5500 1.4 GHz.

📅 Jun 5, 2026 · 📍 Nationwide
Recalling company
Philips North America Llc
Product
Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
Reason
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Jun 5, 2026
Reported
Jul 1, 2026
Distribution
US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.
Status
Ongoing
Quantity
909 units
Lot / codes
Model Number: 867232. UDI-DI: (01)00884838102842. Software Version Number: 4.0.0.0. Serial Numbers: US001A002F, US001A00AW, US001A00C9, US001A00EW, US001A00F0, US001A00GM, US001A00N4, US001A00PB, US001A00PZ, US001A00R1, US001A00UE, US001A00VU, US001A00Z4, US001A013T, US001A018A, US001A018R.
Company location
Cambridge, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-2579-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2579-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.