GE Medical Systems SCS recalls GE Healthcare AW Server 3.2 ext.6.5
- Recalling company
- GE Medical Systems SCS
- Product
- GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
- Reason
- GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- May 8, 2026
- Reported
- Jul 1, 2026
- Distribution
- Worldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Morocco, Nepal, Netherlands, New Zealand, North Macedonia, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, United Arab Emirates, United Kingdom, Vietnam.
- Status
- Ongoing
- Quantity
- 340 servers
- Lot / codes
- REF# (01)00840682102384(10)AWS03D02E6D5, GTIN Number 00840682102384
- Company location
- Buc, N/A
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2553-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2553-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.