Home › FDA recalls › Hillrom VOLARA System P.CIRCUIT KIT AC
CLASS I · SeriousFDA

Baxter Healthcare Corporation recalls Hillrom VOLARA System P.CIRCUIT KIT AC

📅 May 21, 2026 · 📍 Nationwide
Recalling company
Baxter Healthcare Corporation
Product
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
May 21, 2026
Reported
Jul 8, 2026
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Status
Ongoing
Quantity
16187 units
Lot / codes
UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
Company location
Deerfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-2550-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2550-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.