Buy me a coffee
Home › FDA recalls › Off-Axis Comprehensive Shoulder System Small
CLASS II · ModerateFDA

Zimmer, Inc. recalls Off-Axis Comprehensive Shoulder System Small

📅 May 20, 2026 · 📍 Nationwide
Recalling company
Zimmer, Inc.
Product
Off-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240
Reason
Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
May 20, 2026
Reported
Jul 1, 2026
Distribution
US Nationwide distribution.
Status
Ongoing
Quantity
474 units
Lot / codes
Lot Code: Model No. 110040240 ; UDI-DI (01)00889024681965(17)350601(10)099457 ; Lot Number 099457 Model No. 110040240 ; UDI-DI (01)00889024681965(17)350712(10)099458 ; Lot Number 099458
Company location
Warsaw, IN
Recall type
Voluntary: Firm initiated
Recall number
Z-2527-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2527-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.