Intuitive Surgical, Inc. recalls Da Vinci ASSY
- Recalling company
- Intuitive Surgical, Inc.
- Product
- Da Vinci ASSY, PSS, SP1098, Part Number: 380601
- Reason
- Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Apr 22, 2026
- Reported
- Jul 1, 2026
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, GA, OH, PA, NY, MN, TN, OR, IL, MD, NJ, AL, NV, TX, FL, LA, CT, NC, AZ, IA, MO, WA, MI, WV, WI, KY, KS, VA, OK, MA, IN, CO, SC, SD, MS, NH, UT, ME and the countries of Belgium, Denmark, Sweden, France, Japan, Germany, Israel, Switzerland, South Korea, Czech Republic, Austria, United Kingdom, Italy, Spain, Ireland, Saudi Arabia, Greece, Taiwan, Australia, China, Turkey, Poland, United Arab Emirates, Hong Kong, Romania.
- Status
- Ongoing
- Quantity
- 454
- Lot / codes
- UDI: 00886874114605. Serial Numbers: 605408, 606848, 612328, 628399, 609353, 614073, 617794, 616769, 10008062, 635659, 631694, 638771, 641430, 694998, 645433, 643209, 954097, 656203, 764644, 656319, 946098, 667190, 666480, 655562, 664662, 680530, 679266, 683974, 674171, 676962, 682694, 686777, 701248, 696877, 688725, 688225, 697922, 700316, 698277, 702699, 709536, 701751, 717804, 705366, 704731, 724178, 703152, 722803, 693262, 736468, 736338, 740359, 742366, 732233, 696089, 724835, 716520, 709987, 728391, 751529, 731765, 672743, 766366, 711517, 735003, 735252, 746457, 746623, 718940, 721722, 691190, 744094, 750174, 738718, 10211920, 759253, 761448, 752536, 812651, 754935, 790597, 811147, 793019, 761000, 812650, 919988, 919987, 913011, 913012, 926423, 759791, 926422, 938794, 771653, 938795, 766365, 963841, 797872, 769736, 794437, 970568, 970562, 977045, 10003059, 957781, 973824, 804588, 10009706, 936269, 10003509, 930240, 936268, 800142, 10003510, 982782, 967667, 10009707, 973355, 944585, 10065786, 935214, 10020729, 10012963, 10020705, 10031145, 10031144, 973825, 10012962, 10036960, 963856, 10711066, 10035476, 795038, 10043635, 944583, 10072852, 10072853, 966079, 10069899, 10076490, 10087446, 10054478, 10080929, 954096, 10130089, 10165287, 942811, 10180162, 10172589, 10103483, 10103484, 10056518, 10092007, 10246370, 10209961, 10183740, 10232634, 504007, 10231116, 10211880, 10194911, 10245457, 10112470, 10189443, 977044, 10238314, 10644376, 10036961, 10274101, 10112471, 1023831
- Company location
- Sunnyvale, CA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2521-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2521-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.