Beta Bionics, Inc. recalls iLet Bionic Pancreas
- Recalling company
- Beta Bionics, Inc.
- Product
- iLet Bionic Pancreas, REF: BB1001
- Reason
- The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Sep 23, 2025
- Reported
- Jun 24, 2026
- Distribution
- US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL
- Status
- Ongoing
- Quantity
- 4580
- Lot / codes
- iLet ACE Pump firmware BOM000072 Rev C (version 3.3.6) and BOM000072 Rev B (version 3.3.4)
- Company location
- Irvine, CA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2513-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2513-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.