Buy me a coffee
Home › FDA recalls › MEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT
CLASS II · ModerateFDA

Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT CYSTECTOMY

📅 May 4, 2026 · 📍 Nationwide
Recalling company
Medline Industries, LP
Product
MEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT CYSTECTOMY, D AND C PACK, MAD D & C, VPH LITHOTOMY MINOR, TRMC- PERI GYN PACK, RICH MINOR LITHOTOMY, etc. (see recall documents for a full list of products)
Reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
May 4, 2026
Reported
Jun 24, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Status
Ongoing
Quantity
32438 units
Lot / codes
Medline Kit Number/SKU DYNJ83242B: UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25HBR321; Medline Kit Number/SKU DYNJ83242B: UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25GBD202; Medline Kit Number/SKU DYNJ83242B: UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25FBK813; Medline Kit Number/SKU DYNJ83242B: UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25EBG207; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 23FBK499; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 23EBV071; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22KBK089; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22IBV707; Medline Kit Number/SKU DYNJ61059B: UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22HBC310; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 26DMG168; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 26CMD637; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25KMC849; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25JMA672; Medline Kit Number/SKU DYNJ61059C: UDI/DI 10195327451653
Company location
Northfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-2506-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2506-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.