Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY
- Recalling company
- Medline Industries, LP
- Product
- MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR GYN PACK, FS LAP TUBAL LIGATION, HILLSDALE D&C / HYSTEROSCOPY, LAVH NEW BRAUNFELS,TVH PACK, etc (see recall documents for a full list of affected products)
- Reason
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- May 4, 2026
- Reported
- Jun 24, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
- Status
- Ongoing
- Quantity
- 37324 units
- Lot / codes
- Medline Kit Number/SKU DYNJ61063A: UDI/DI 10193489640458 (each), 40193489640459 (case), Lot Number 21KBV571; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23EBU351; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23DBQ190; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23BBE364; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 22IBW094; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26CMK666; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26BMG709; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26AMI644; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25KMD133; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25JMA568; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25IMF717; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25HMJ595; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25FMH149; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684
- Company location
- Northfield, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2505-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2505-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.