D.O.R.C. Dutch Opthalmic Research Center Intl B.V. recalls TDC VELOCE"
- Recalling company
- D.O.R.C. Dutch Opthalmic Research Center Intl B.V.
- Product
- TDC VELOCE", 23G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT23, 9311.23G01, 9311.23G02, 9311.23K01, 9311.23K02, 9311.23K03, 9623.G0201, 9623.G0202, 9623.K0201, 9311.23G01-00;
- Reason
- An increase in incidence in complaints of a loose outer knife and reports of the outer knife becoming loose during use.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Apr 30, 2026
- Reported
- Jun 17, 2026
- Distribution
- Worldwide - US Nationwide and the countries of AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CZECHIA, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, HONG KONG, IRELAND, ITALY, JAPAN, NETHERLANDS, NEW ZEALAND, POLAND, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, TAIWAN, TURKEY.
- Status
- Ongoing
- Quantity
- 314 boxes, totaling 1884 units
- Lot / codes
- 1. Article Number: 9268.VIT23; UDI-DI: 08717872031866; 2. Article Number: 9311.23G01; UDI-DI: 08717872035109; 3. Article Number: 9311.23G02; UDI-DI: 08717872035338; 4. Article Number: 9311.23K01; UDI-DI: N/A; 5. Article Number: 9311.23K02; UDI-DI: N/A; 6. Article Number: 9311.23K03; UDI-DI: N/A; 7. Article Number: 9623.G0201; UDI-DI: N/A; 8. Article Number: 9623.G0202; UDI-DI: N/A; 9. Article Number: 9623.K0201; UDI-DI: N/A; 10. Article Number: 9311.23G01-00; UDI-DI: 08717872035307; Lot Numbers: 100637001, 57682-*-*-1, 64312-*-*-1, 69017-*-*-1, 73237-*-*-1, 74225-*-*-1, 75735-*-*-1, 79460-*-*-1, 80338-*-*-1, 80963-*-*-1, 81335-*-*-1, 81679-*-*-1, 82213-*-*-1, 82325-*-*-1, 82742-*-*-1, 82779-*-*-1, 83584-*-*-1, 84580-*-*-1, 85451001, 86069001, 86798001, 86799001, 87205001, 88231001, 91957001, 92629001, 92630001, 93336001, 94351001, 95436001, 95462001, 96097001, 96123001, 97279001, 97789001, 98643001, 99130001, 99185001, 99697001;
- Company location
- Zuidland, N/A
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2322-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2322-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.