Home › FDA recalls › Plum Solo Precision IV Pump
CLASS II · ModerateFDA

ICU Medical, Inc. recalls Plum Solo Precision IV Pump

📅 Apr 29, 2026 · 📍 Nationwide
Recalling company
ICU Medical, Inc.
Product
Plum Solo Precision IV Pump, 40001-0401
Reason
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Apr 29, 2026
Reported
Jun 17, 2026
Distribution
Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.
Status
Ongoing
Quantity
13,613
Lot / codes
UDI-DI M335400011 Plum Solo Precision IV Pump v1.0.3
Company location
Lake Forest, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-2311-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2311-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.