Home › FDA recalls › CASE v7.0 Cardiac Testing System
CLASS II · ModerateFDA

GE Medical Systems Information Technologies Inc recalls CASE v7.0 Cardiac Testing System

📅 Apr 24, 2026 · 📍 Nationwide
Recalling company
GE Medical Systems Information Technologies Inc
Product
CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3
Reason
Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Apr 24, 2026
Reported
Jun 10, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Bangladesh, Cambodia, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, South Korea, Lebanon, Libya, Lithuania, Malaysia, Myanmar, Netherlands, New Zealand, Norway, Philippines, Poland, Qatar, Romania, Saudi Arabia, Senegal, Singapore, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom.
Status
Ongoing
Quantity
1835
Lot / codes
UDI-DI 00195278617644 Serial numbers SU325232838SA SU325032272SA SU325293027SA SU325222794SA SU325122553SA SU325363223SA SU325483517SA SU325152591SA SU325363222SA SU325483514SA SU325272955SA SU325222809SA SU325092421SA SU325272941SA SU325272944SA SU325393292SA SU325393293SA SU325393296SA SU325443415SA SU325303053SA SU325353201SA SU324512241SA SU325032260SA SU325433377SA SU325242899SA SU325242900SA SU325192689SA SU325172624SA SU325172626SA SU325172630SA SU325333173SA SU325383266SA SU324502188SA SU325112523SA SU325112524SA SU325192719SA SU325313091SA SU325403321SA SU325323101SA SU325323103SA SU324502193SA SU324502203SA SU324502209SA SU324502211SA SU325272960SA SU325272962SA SU325252906SA SU325433376SA SU325473497SA SU324502191SA SU324502204SA SU324502206SA SU324502207SA SU325293013SA SU325293014SA SU325293015SA SU325293016SA SU325383265SA SU324482168SA SU325122538SA SU325122540SA SU325192691SA SU325192692SA SU325192708SA SU324492172SA SU325172632SA SU325182669SA SU325323135SA SU325323137SA SU325503570SA SU326033668SA SU325172613SA SU325052318SA SU325373241SA SU325443399SA SU325222802SA SU325222803SA SU325222805SA SU325293025SA SU325272954SA SU324512242SA SU325132576SA SU325172612SA SU325172614SA SU325172640SA SU325172641SA SU325192693SA SU325252910SA SU325092412SA SU325333171SA SU325122543SA SU325182687SA SU325242868SA SU325242870SA SU325473487SA SU326023667SA SU325062342SA SU325473493SA SU325252909SA SU325282987SA SU325293020SA SU325473494SA SU325473495SA SU326033696SA SU32603
Company location
Milwaukee, WI
Recall type
Voluntary: Firm initiated
Recall number
Z-2279-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2279-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.