Oculus Technologies of Mexico, S.A. de C.V. recalls PhaseOne Antimicrobial Solution
- Recalling company
- Oculus Technologies of Mexico, S.A. de C.V.
- Product
- PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
- Reason
- The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Apr 27, 2026
- Reported
- Jun 10, 2026
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.
- Status
- Ongoing
- Quantity
- 5904 units
- Lot / codes
- UDI:00860009268616 Lot: 25G263
- Company location
- Zapopan
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2275-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2275-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.