PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura system
- Recalling company
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039);
- Reason
- It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Apr 30, 2026
- Reported
- Jun 3, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
- Status
- Ongoing
- Lot / codes
- 1. Allura Xper FD10; Model: 722003; UDI-DI: N/A; Serial Numbers: 445, 1555, 1114, 1529, 472, 1349, 1594, 1311, 1472, 1187, 1700, 1031, 912, 913, 940, 285, 1637, 5, 797, 639, 452, 1585, 535, 1338, 1641, 1916, 614, 605, 492, 1612, 505, 239, 1216, 1169, 1118, 115, 519, 356, 715, 423, 1399, 789, 1509, 1016, 875, 1923, 313, 314, 1919, 1458, 369, 1176, 1230, 1925, 339, 1325, 1633, 596, 1784, 464, 796, 1928, 1025, 292, 1669, 1697, 708, 1, 1760, 1452, 346, 961, 1172, 354, 416, 363, 582, 161, 569, 922; 2. Allura Xper FD10; Model: 722010; UDI: (01)00884838059030(21)1123, (01)00884838059030(21)932, (01)00884838059030(21)239, (01)00884838059030(21)360, (01)00884838059030(21)500, (01)00884838059030(21)794, (01)00884838059030(21)939, (01)00884838059030(21)921, (01)00884838059030(21)286, (01)00884838059030(21)1440, (01)00884838059030(21)1144, (01)00884838059030(21)521, (01)00884838059030(21)33, (01)00884838059030(21)107, (01)00884838059030(21)1025, (01)00884838059030(21)668, (01)00884838059030(21)656, (01)00884838059030(21)1261, (01)00884838059030(21)1170, (01)00884838059030(21)585, (01)00884838059030(21)1128, (01)00884838059030(21)882, (01)00884838059030(21)1230, (01)00884838059030(21)280, (01)00884838059030(21)488, (01)00884838059030(21)787, (01)00884838059030(21)968, (01)00884838059030(21)890, (01)00884838059030(21)780, (01)00884838059030(21)190, (01)00884838059030(21)1181, (01)00884838059030(21)694, (01)00884838059030(21)1279, (01)00884838059030(21)344, (01)00884838059030(21)1149, (01)0
- Company location
- Best
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2259-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2259-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.