Buy me a coffee
Home › FDA recalls › Azurion system
CLASS II · ModerateFDA

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion system

📅 Apr 30, 2026 · 📍 Nationwide
Recalling company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222), Azurion 5 M12 (722227), Azurion 5 M20 (722228), Azurion 7 B12 (722067, 722225), Azurion 7 B20 (722068, 722226), Azurion 7 M12 (722078, 722223), Azurion 7 M20 (722079, 722224).
Reason
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Apr 30, 2026
Reported
Jun 3, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Status
Ongoing
Quantity
3069 units (872 US, 2224 OUS)
Lot / codes
1. Azurion 3 M12; Model: 722063; UDI: (01)00884838085275(21)122, (01)00884838085275(21)64, (01)00884838085275(21)56, (01)00884838085275(21)131, (01)00884838085275(21)84, (01)00884838085275(21)132, (01)00884838085275(21)98, (01)00884838085275(21)108, (01)00884838085275(21)77, (01)00884838085275(21)82, (01)00884838085275(21)47, (01)00884838085275(21)80, (01)00884838085275(21)146, (01)00884838085275(21)102, (01)00884838085275(21)117, (01)00884838085275(21)2, (01)00884838085275(21)149, (01)00884838085275(21)99, (01)00884838085275(21)44, (01)00884838085275(21)3, (01)00884838085275(21)28, (01)00884838085275(21)144, (01)00884838085275(21)62, (01)00884838085275(21)119, (01)00884838085275(21)61, (01)00884838085275(21)22, (01)00884838085275(21)29, (01)00884838085275(21)40, (01)00884838085275(21)70, (01)00884838085275(21)30, (01)00884838085275(21)25, (01)00884838085275(21)19, (01)00884838085275(21)120, (01)00884838085275(21)121, (01)00884838085275(21)110, (01)00884838085275(21)16, (01)00884838085275(21)14, (01)00884838085275(21)17, (01)00884838085275(21)53, (01)00884838085275(21)103, (01)00884838085275(21)101, (01)00884838085275(21)143, (01)00884838085275(21)41, (01)00884838085275(21)104, (01)00884838085275(21)116, (01)00884838085275(21)151, (01)00884838085275(21)93, (01)00884838085275(21)141, (01)00884838085275(21)115, (01)00884838085275(21)73, (01)00884838085275(21)150, (01)00884838085275(21)128, (01)00884838085275(21)105, (01)00884838085275(21)4, (01)00884838085275(21)50, (01)0088483
Company location
Best
Recall type
Voluntary: Firm initiated
Recall number
Z-2258-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2258-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.