GE Medical Systems Information Technologies Inc recalls GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L
- Recalling company
- GE Medical Systems Information Technologies Inc
- Product
- GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason
- GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Apr 24, 2026
- Reported
- Jun 3, 2026
- Distribution
- Worldwide distribution.
- Status
- Ongoing
- Quantity
- 2366 units
- Lot / codes
- No UDI, Serial Number or Sales Order Number: KSA7536146, KSA7536158, KSA7536159, KSA7550156, KSA7550158, KSA7550159, KSA7667317, KSA7754389, KSA7769597, KSA7769610, KSA7769612, KSA7773420, KSA7773431, KSA7773440, KSA7120704, KSA7120767, KSA6964886, KSA6808842, KSA6808852, KSA6808866, KSA6808868, KSA6808881, KSA6808883, KSA6808912, KSA6808928, KSA6866974, KSA6664007, KSA6866963, KSA7120766, KSA7162586, KSA6623197, KSA6623216, KSA6623223, KSA6623226, KSA6623239, KSA6623248, KSA6623270, KSA6623278, KSA7391460, KSA6477094, KSA6477095, KSA6477135, KSA6477155, KSA6477159, KSA6477206, KSA6954245, KSA6954269, KSA6954276, KSA6954277, KSA6954284, KSA6954291, KSA6477096, KSA6477154, KSA7702196, KSA7702198, KSA7702199, KSA7702208, KSA7702222, KSA6623191, KSA6623194, KSA6623237, KSA6623251, KSA6623267, KSA6626626, KSA6626632, KSA6626639, KSA6626666, KSA6626670, KSA6626716, KSA6536442, KSA6536443, KSA6536454, KSA6536460, KSA6536469, KSA6536484, KSA6536485, KSA6536490, KSA6536496, KSA6536504, KSA6536526, KSA6536528, KSA6536532, KSA7120699, KSA7120728, KSA7120764, KSA7120773, KSA7162570, KSA7162593, KSA7162610, KSA7162622, KSA7162640, KSA6623198, KSA6623202, KSA6623232, KSA6623236, KSA6623275, KSA6623280, KSA6626622, KSA6626625, KSA6626635, KSA6626636, KSA6626648, KSA6626655, KSA6626659, KSA6626661, KSA6536520, KSA6536525, KSA7304431, KSA6866921, KSA6866926, KSA6866927, KSA6866983, KSA6808882, KSA6649195, KSA6649250, KSA7162575, KSA7162588, KSA7162608, KSA7162614, KSA7162656, KSA6899015, KS
- Company location
- Waukesha, WI
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2246-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2246-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.