Abiomed, Inc. recalls Automated Impella Controller

📅 Jan 17, 2011 · 📍 Nationwide
Recalling company
Abiomed, Inc.
Product
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US.
Reason
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Jan 17, 2011
Reported
Jun 3, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Denmark, France, Germany, Italy, Kuwait, Netherlands, Norway, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom.
Status
Ongoing
Quantity
625 units
Lot / codes
1. Product Code: 0042-0000-CA. GTIN: 00813502011272. Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715. 2. Product Code: 0042-0000-EU. GTIN: 00813502011289. Serial Numbers: IC1217, IC1265, IC1275, IC1276, IC1288, IC1284, IC1301, IC1302, IC1304, IC1305, IC1344, IC1341, IC1343, IC1340, IC1345, IC1350, IC1347, IC1349, IC1383, IC1381, IC1384, IC1386, IC1393, IC1396, IC1397, IC1395, IC1413, IC1415, IC1411, IC1416, IC1490, IC1488, IC1522, IC1523, IC1524, IC1525, IC1547, IC1549, IC1546, IC1548, IC1578, IC1577, IC1579, IC1580, IC1594, IC1584, IC1590, IC1592, IC1591, IC1585, IC1588, IC1653, IC1676, IC1685, IC1684, IC1677, IC1678, IC1699, IC1700. 3. Product Code: 0042-0000-UK. GTIN: 00813502011296. Serial Numbers: IC1387, IC1389, IC1388, IC1519, IC1542, IC1545, IC1581, IC1582, IC1583, IC1573. 4. Product Code: 0042-0000-US. GTIN: 00813502010022. Serial Numbers: IC1004, IC1002, IC1001, IC1006, IC1007, IC1008, IC1014, IC1015, IC1017, IC1019, IC1024, IC1025, IC1028, IC1029, IC1026, IC1027, IC1033, IC1035, IC1031, IC1032, IC1036, IC1037, IC1038, IC1039, IC1041, IC1042, IC1044, IC1049, IC1043, IC1046, IC1053, IC1055, IC1052, IC1040, IC1030, IC1056, IC1047, IC1051, IC1050, IC1060, IC1061, IC1062, IC1064, IC1065, IC1066, IC1067, IC1074, IC1070, IC1071, IC1072, IC1073, IC1068, IC1069, IC1058, IC1075, IC1076, IC1077, IC1078, IC1079, IC1084, IC1085, IC1083, IC1080, IC1088, IC1086, IC1087, IC1081, IC1082, IC1100, IC1099, IC1093, IC
Company location
Danvers, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-2238-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-2238-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.