PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura system
- Recalling company
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039), Cardio Vascular-Allura (722400);
- Reason
- The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Apr 28, 2026
- Reported
- May 27, 2026
- Distribution
- Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen;
- Status
- Ongoing
- Quantity
- 3314 units (816 US, 2498 OUS)
- Lot / codes
- Code Information: Allura: 1. Allura Xper FD10C; Model: 722001; UDI-DI: N/A; Serial Numbers: 170, 242, 267, 142, 206, 316, 9, 306, 103, 33, 220; 2. Allura Xper FD10F; Model: 722002; UDI-DI: N/A; Serial Numbers: 0, 81, 47; 3. Allura Xper FD10; Model: 722003; UDI-DI: N/A; Serial Numbers: 208, 392, 109, 45, 1953, 1471, 1551, 485, 152, 928, 501, 90, 1435, 81, 939, 160, 472, 1349, 439, 1958, 1525, 72, 1187, 912, 368, 328A, 37, 1286, 1223, 487, 1061, 833, 688, 317, 157, 639, 375, 420, 225, 1338, 614, 214, 230, 736, 64, 1090, 1094, 1932, 1531, 421, 984, 373, 313, 918, 602, 374, 750, 1934, 1751, 1661, 961, 791, 454, 868, 416, 1329, 241; 4. Allura Xper FD10; Model: 722010; UDI: (01)00884838059030(21)921, (01)00884838059030(21)1440, (01)00884838059030(21)1328, (01)00884838059030(21)1292, (01)00884838059030(21)1428, (01)00884838059030(21)1396, (01)00884838059030(21)298, (01)00884838059030(21)1432, (01)00884838059030(21)1425, (01)00884838059030(21)1398, (01)00884838059030(21)1397, (01)00884838059030(21)1391, (01)00884838059030(21)1395, (01)00884838059030(21)314, (01)00884838059030(21)1297, (01)00884838059030(21)946, (01)00884838059030(21)1424, (01)00884838059030(21)1244, (01)00884838059030(21)1122, (01)00884838059030(21)1198, (01)00884838059030(21)1421, (01)00884838059030(21)1423, (01)00884838059030(21)1438, (01)00884838059030(21)697, (01)00884838059030(21)1192, (01)00884838059030(21)1326, (01)00884838059030(21)402, (01)00884838059030(21)1148, (01)00884838059030(21)1439, (01)0088483805903
- Company location
- Best
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2228-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2228-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.