Fresenius Kabi USA, LLC recalls Ivenix Infusion System
- Recalling company
- Fresenius Kabi USA, LLC
- Product
- Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
- Reason
- Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- May 6, 2026
- Reported
- Jun 10, 2026
- Distribution
- Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
- Status
- Ongoing
- Quantity
- 18,444
- Lot / codes
- Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:
- Company location
- North Andover, MA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-2224-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-2224-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.